Technology, Reliability, Integrity
HomeNewsAbout UsServiceRepair TagLocationsProductsElect AdaptersSales
Exhibitions MFC Tutorial Industry Links Contact Us

800-MFC-HELP

Technology Reliability Integrity Home News About Us Service Repair Tag Locations Products Elect Adapters Sales Exhibitions MFC Tutorial Industry Links Contact Us
NIST

Sample COC

coastal instruments

Mass Flow Controller Specialists

Post Office Box 699 - Burgaw, NC 28425 - 910.259.4485

Certificate of Calibration

Certificate #:
NC 00000000

The following device has been calibrated in accord with ANSI Z540 and ISO/IEC-17025 using instruments traceable to the National Institute of Standards and Technology.

Customer Name
1234 Main St.
Anywhere, USA

STP Conditions 0.0 Deg. C @ 760 mmHG
Calibration Temperature 22.2° C ± 1° C
He pre-ship leak rate 0.9 x 10- 10
Model:
Serial Number:
Range:
Gas:
Inlet Pressure:
Outlet Pressure:
Positioning:
Calibration Gas:
Calibration Date:

MKS 1479A13CR1AM
750460
1000 sccm
N2
10 PSIG
0.6 TORR
VIU
N2, N2 CF = 1
6-2-2005

Initial Inspection
No visible contamination

Services
Disassembled
Purged sensor with methyl alcohol and N2
Cleaned parts with Alcohol and DI water
Calibrated using Coastal tranfer standards. IAW Document 005, "MFC Calibrations"
Assembled with Metal seals

Parts Replaced
No parts required

Comments
We appreciate your business!
PO #: 999999

Post Cal Volts % Calc. Flow (sccm)
F.S Dev.
  0.012
1.0
2.0
3.0
4.0
5.0
0
20
40
60
80
100
No Flow
202.95
400.848
596.901
794.775
999.5
0.24%
0.29%
0.08%
-0.31%
-0.52%
-0.05%
Pre Cal   0 No Flow 0.00%
  1.0
2.0
3.0
4.0
5.0
20
40
60
80
100


No Pre Cal
data taken


 

Uncertainity of measurement = 0.37% of F.S. (95% confidence level,k = 2)

%F.S. Dev. Calculated Target * 100
     Full Scale

 

This report not to be reproduced, except in full, without written authorization by Coastal Instruments, Inc.